Kewtech KT66DL Digital 12-in-1 & EVSE Adapter
Includes: KT66DL, KEWEVSE EVSE Adapter, ACC064SP G7 Test probe, with remote test button, ACC065 G7 Test lead set, Kamp12 Mains lead, USB cable, Test lead pouch, Batteries, Soft carry case, Padded strap, Instruction Manual, Calibration Certificate.
Our pages are encrypted and protected from cyber-attack.
Payment Card Industry Data Security Standard (PCI DSS) Compliant.
TLS (SSL) - all data/information is secure from interception.
The Kewtech KT66DL and KEWEVSE testing adapter for the testing of EV charging points provides a comprehensive kit for those with electrical vehicle charging points. Whether these are installed with Type B RCDs or with 6mA residual direct current devices and in either TN or TT configurations, this kit gives you the flexibility to test for electrical safety as well as EVSE functionality.
The KT66DL has an easy-to-use single rotary dial with four function softkeys and a colour LCD 3.5” dot matrix display. It also incorporates Anti-Trip Technology (ATT) which electronically bypasses RCDs when performing loop impedance tests. This saves time and money by not having to take the RCD out of the circuit during testing and is a safer procedure to follow.
The KT66DL Multi-function tester has eight test functions in one tester.
1. Continuity test.
2. Insulation resistance test.
3. Loop impedance test (High current measurement, No trip (ATT) measurement).
4. Prospective short circuit current test (On Loop impedance function).
5. RCD test.
6. Voltage test.
7. Phase rotation test.
8. Earth test.
The KT66DL incorporates Anti Trip Technology (ATT) which electronically by-passes RCDs when performing loop impedance tests. This saves time and money by not having to take the RCD out of the circuit during testing and is a safer procedure to follow.
With the ATT function enabled, a test of 15mA or less is applied between line & earth. It enables loop impedance measurements without tripping RCDs rated at 30mA and above. ATT supports measurements using three wires: Line, Earth, and Neutral and also two
wires: Line and Earth.
FEATURES
18th Edition Functions.
Phase Rotation.
Insulation test: 4 test voltages, resistance bar graph & DAR / PI functionality.
Two continuity tests 200 mA & 15 mA.
0.001 Ω res high current loop.
55V Loops for 110V CTE transformers.
Earth resistance 3 & 2 wire options.
Advanced memory with PAT test download.
Two wire low current loop.
EVSE loop option.
RCD - Tests Type AC, ACS, A, AS, F, B, BS & EV.
EVSE 6mA DC ramp and tripping time test.
PAT function including download.
Continuity | ||||||
---|---|---|---|---|---|---|
Range | 20.00/200.0/2000Ω (auto-ranging). | |||||
Display range | 0.00 - 20.99Ω / 16.0 - 209.9Ω / 160 - 2099Ω. | |||||
Measuring range | 0 – 2000Ω. | |||||
Accuracy | ±2.0%rdg±8dgt. | |||||
Open-circuit voltage | 7 – 14V DC. | |||||
Test current | 200mA test: 200mA or more (2Ω or less) 15mA test: 15mA±3mA (short-circuit). |
|||||
Enabled if pre-set NULL value is 9Ω or less. Buzzer: Buzzer sounds when measured resistance is 2Ω or less. | ||||||
Volts | ||||||
Range | 300.0/600V (Auto-ranging) | |||||
Display range | Voltage: 2.0 – 314.9V, 240 – 629V Frequency: 40.0 – 70.0Hz (displayed at 2V or higher). |
|||||
Measuring range | Voltage: 2 – 600V | |||||
Accuracy | Voltage: ±2%rdg±4dgt Frequency: ±0.5%rdg±2dgt. |
|||||
Uc (RCD) | ||||||
Mains voltage range | 100-260V. | |||||
Range | 100V. | |||||
Display range | 0.0 - 104.9V. | |||||
Measuring range | 0 - 100V. | |||||
Accuracy | +5%to+15%rdg ±8dgt 230V. | |||||
Test current | 50 % or less of IΔn. | |||||
Earth | ||||||
Precise measurement. | Simplified measurement. | |||||
Range | 20.00 / 200.0 / 2000Ω (auto-ranging). | |||||
Accuracy | 20Ω range; ±2%rdg±0.08Ω The other ranges: ±2%rdg±3dgt (Aux. earth resistance: 100Ω). |
±2%rdg±0.08Ω The other ranges: ±2%rdg±3dgt. |
||||
Output current | 20Ω range: approx. 3mA, 200v range: approx. 1.7mA 2000Ω range: approx. 0.7mA Frequency: 825Hz. |
|||||
Phase Rotation | ||||||
Measuring Range | 48 – 600V / 45 - 65Hz. | |||||
Criteria of judgement | Correct sequence: Clockwise symbol and “1, 2, 3” are displayed Reversed sequence: Counter-clockwise symbol and “3, 2, 1” are displayed. |
|||||
Insulation | ||||||
Rated voltage | 100V | 250V | 500V | 1000V | SPD (Varistor) 1000V | |
Range | 2.000/20.00/200.0MΩ auto-ranging |
20.00/200.0/1000MΩ auto-ranging |
20.00/200.0/2000MΩ | 0 -1000V (increases in 1V steps) | ||
Display range | 0.000 - 2.099MΩ | 0.00 - 20.99MΩ | 0.00 - 20.99MΩ | 1049V | ||
1.60 - 20.99MΩ | 16.0 - 209.9MΩ | 16.0 - 209.9MΩ | ||||
16.0 - 209.9MΩ | 160 - 1049MΩ | 160 - 2099MΩ | ||||
Measuring range | 0 - 200MΩ | 0 - 1000MΩ | 0 - 2000MΩ | 0 - 1049V | ||
Accuracy | 2.000MΩ range: ±2%rdg±6dgt | 20.00MΩ range: ±(2%rdg+6dgt), 200.0MΩ range: ±(2%rdg+6dgt) | ±5%rdg±5dgt | |||
20.00MΩ range: ±2%rdg±6dgt | 1000MΩ range: ±(5%rdg+6dgt) | 2000MΩ range: ±(5%rdg+6dgt) | ||||
200MΩ range: ±5%rdg±6dgt | ||||||
Rated measurement current | 1.0 - 1.2mA at 100kΩ |
1.0 - 1.2mA at 250kΩ |
1.0 - 1.2mA at 500kΩ |
1.0 - 1.2mA at 1MΩ |
1.0 mA | |
Loop | ||||||
Function | LOOP ATT | LOOP HIGH | ||||
3-Wire L-PE | 2-Wire L-PE | L-PE0.01ΩRes | L-PE0.001ΩRes | L-N/L-L | ||
Mains input voltage range | 100 – 260V 50/60Hz (L-N < 20Ω) |
48 – 260V 50/60Hz | 100 - 260V 50/60Hz | 48 - 260V 50/60Hz | ||
Range | LOOP | 20.00/200.0/2000Ω (auto-ranging) | 20.00/200.0/2000Ω | 2.000Ω | 20.00Ω | |
PFC/PSC | 2000A/20kA | 2000A/20kA (PFC only) | 2000A/50kA (PFC only) | 2000A/20kA (PSC only) | 2000A/20kA (PSC only) | |
Display range | LOOP | 0.00 – 20.99Ω 21.0 – 209.9Ω 210 – 2099Ω |
0.00 –20.99Ω 21.0 –209.9Ω 210 –2099Ω |
0.00 - 20.99Ω 21.0 - 209.9Ω 210 –2099Ω |
0.000 - 2.099Ω | 0.00-20.99Ω |
PFC/PSC | 0 – 2099A 2.10 – 20.99kA |
0 – 2099A 2.10 – 20.99kA (PFC only) |
0 – 2099A 2.10 – 50.49kA (PFC only) |
0 – 2099A 2.10 – 20.99kA (PFC only) |
0 – 2099A 2.10 – 20.99kA (PSC only) |
|
Meas. range | LOOP | 0 – 2000Ω | 0 – 2000Ω | 0 – 2000Ω | 0 – 2Ω | 0 – 20Ω |
Accuracy | LOOP | ±(3%rdg+6dgt) | ±(3%rdg+10dgt) | ±(3%rdg+4dgt) | ±(3%rdg+25mΩ) | ±(3%rdg+4dgt) |
PFC/PSC | Depending on the accuracies of voltage and LOOP measurements. | |||||
Test current @ 230V | L-N:6A/60ms N-PE:10mA (5.3Hz) |
L-PE: 15mA | 20Ω: 6A/20ms, 200Ω: 0.5A/20ms 2000Ω: 15mA/500ms |
25A/20ms | 6A/20ms | |
RCD | ||||||
Mains input voltage range: 100V – 260V 50/ 60Hz. For Type AC and A RCDs rated at 100mA or higher: 190 – 260V. | ||||||
Mode | Rated residual operating current (mA)(IΔn) | Test current | Duration | |||
Current (mA) rms | Accuracy @230V | Meas. time | Accuracy | |||
x 1/2 | AC | G: 10/30/100/300/500/1000 S: 100/300/500 | IΔn×1/2 | -8% to-2% VAR:-10% to 0% | 2000ms | Trip Time ±(1%+2ms) Measuring time ±3% of FS |
A/F | G: 10/30/100/300/500 AS:100/300/500 | IΔn×1/2 | -10% to 0% | |||
B | G: 10/30/100/300 S:100/300 | IΔn×1/2 | -10% to 0% | |||
x 1 | AC | G:10/30/100/300/500/1000 S: 100/300/500 | IΔn | +2% to +8% VAR:0% to +10% | G:550ms S:1000ms |
|
A/F | G: 10/30/100/300/500 AS:100/300/500 | IΔn×1.4 (10mA: IΔn×2) | 0% to +10% | |||
B | G: 10/30/100/300 S:100/300 | IΔn×2 | 0% to +10% | |||
6 mA DC | IΔn | 0% to +10% | 10.5s | |||
x 5 | AC | G: 10/30/100 S: 100 | IΔn×5 | +2% to +8% VAR:0% to +10% | 410ms | |
A/F | G: 10/30/100 AS: 100 | IΔn×5×1.4 | 0% to +10% | |||
B | 10/30 | IΔn×2×5 | 0% to +10% | |||
Ramp 20 to 110% |
AC | 10/30/100/300/500 S: 100/300/500 | IΔn | -4% to +4% | by 10% G: 300ms S: 500ms |
Measuring time ±3% of FS |
A/F | G: 10/30/100/300/500 AS:100/300/500 | IΔn×1.4 (10mA:IΔn×2) | -10% to +10% | |||
B | G: 10/30/100/300 S: 100/300 | IΔn×2 | -10% to +10% | by 2% /150ms | ||
Ramp EV 30 to 100% | 6 mA DC | IΔn | -10% to +10% | by 2% / 500ms | - | |
General | ||||||
Applicable Standards | IEC 61010-1 CATIII (600V) CATIV (300V) pollution degree 2, IEC 61010 -Tester , -2-030, -2-034, IEC61557-1,2,3,4,5,6,7,10, IEC 60529 IP40, EN 61326-2-2 (EMC). |
|||||
Battery type | Size AA Alkaline battery (LR6) x 8. | |||||
Display | >Colour Dot Matrix LCD 320(W) X 240(H) pixels. | |||||
Operating temperature | -10 to +50°C. | |||||
Dimensions | 136 (L) x 235 (W) x 114 (D) mm. | |||||
Weight | 1300g including batteries (8 x AA / LR6). |
Input voltage | Max. 250V (single-phase), max. 430V (three-phase) 50/60Hz. |
Measurement Category | CAT II 300V. |
Mains socket rating | 250V/10A. |
Mains socket protection | Fuse T10A/250V. |
PP simulation | Open circuit, 13A, 20A, 32A, 63A. |
CP simulation | States A, B, C, D. |
Error simulation | CP error “E”, PE (earth fault) error. |
PE Pre-Test | Yes. |
Test connector type | IEC62196-2 Type 2 male. |
Test cable length | 25cm. |
Working temperature | 0 … +40ºC. |
Storage temperature | -10 … +50ºC. |
Humidity | 0-80% RH. |