The Fluke 9009 includes two independently controlled temperature blocks. The hot block provides temperatures from 50°C to 350°C, while the cold block covers the range –15°C to 110°C.

Each block is controlled by a precision Fluke Calibration temperature controller. The controller uses a precision platinum RTD as a sensor and controls the hot temperature block with a solid state relay (triac) driven heater. A thermal electric device (TED) controls the cold temperature block.

The LED front panel continuously shows the current well temperature. The temperature may be easily set with the control buttons to any desired temperature within the specified range. The calibrator’s multiple fault protection devices insure user and instrument safety and protection.

The 9009 is housed in a tough Pelican case that is both airtight and watertight. It’s a small package weighing only 10 pounds, yet it fits everything you need, including a power cord and four extra inserts. Inserts are available to accommodate sensors of any size from 1/16" (1.6 mm) to 7/16" (11.1 mm).

FEATURES

Two independently controlled temperature blocks.
Rapid heating and cooling.
Interchangeable multiple hole probe sleeves.
Convenient integrated carrying case.
RS-232 interface.
Built in programmable features.
Temperature scan rate control.
Eight set-point memory.
Indoor use only.

Fluke Calibration 9009 Specifications
Hot block Cold block
Range 50°C to 350°C (122°F to 662°F) –15°C to 110°C (5°F to 230°F)
(–8°C [18°F] with hot block at 350°C [662°F]).
Accuracy ±0.6°C ±0.2°C.
Stability ±0.05°C.
Well-to-well uniformity ±0.1°C.
Display resolution 0.1°.
Heating times 30 minutes from 25°C to 350°C 15 minutes from 25°C to 110°C.
Cooling times 40 minutes from 350°C to 100°C 16 minutes from 25°C to –15°C.
Stabilization times 8 minutes.
Well depth 4 in (102 mm).
Removable inserts Two 6.4 mm (1/4 in) and two 4.8 mm (3/16 in) inserts included; see ordering information for other available inserts.
Computer interface RS-232 included with free Interface-it software.
Power 115 V AC (±10 %), 3 A, or
230 V AC (±10 %), 2 A, specify, 50/60 Hz, 280 W.
Dimensions (H x W x D) 178 x 267 x 248 mm (7 x 10.5 x 9.75 in).
Weight 4.5 kg (10 lb).
NIST-traceable calibration Data at 50°C, 100°C, 150°C, 200°C, 250°C, 300°C, and 350°C Data at –8°C, 0°C, 25°C, 50°C, 75°C, 100°C, and 110°C.