- settings_phone Telephone: 02890 830302
- mail belfast@cuthbertsonlaird.co.uk
- EUR €
- GBP £
Fluke 5611A-11-X Secondary Reference Thermistor Probe
Silicone-Bead Probe, 1.5 mm dia x 25.4 mm (0.06 in x 1 in), 0 °C to 100 °C, Lead wire length (Cable) 6 ft.
Our pages are encrypted and protected from cyber-attack.
Payment Card Industry Data Security Standard (PCI DSS) Compliant.
TLS (SSL) - all data/information is secure from interception.
Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “I" (INFO-CON for 1521 or 1522), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs).
All models come without calibration unless calibration is purchased separately.
With a diameter at the tip of just 1.5 mm (0.06 in), the Fluke 5611A's tip has the smallest diameter of any of our secondary reference thermistors and can fit nearly anywhere. Its faster response time, flexible sheath, and silicone coating make the Fluke 5611A great for use in many applications. However, applications involving silicone oil could damage the thermistor and should be avoided.
FEATURES:Silicone-Bead Probe, 1.5 mm dia x 25.4 mm (0.06 in x 1 in), 0 °C to 100 °C, Lead wire length (Cable) 6 ft.
Accredited NVLAP calibration optional.
Resistance | Nominal 10,000 Ω at 25 °C. |
Range | 0 °C to 100 °C. |
Calibration | R vs. T table with 0.1 °C increments, interpolation equation furnished. |
Calibration Uncertainty | Table and equation are accurate to ± 0.01 °C Optional accredited† calibration uncertainties ± 0.006 °C. |
Drift | Better than ± 0.01 °C per year (k=3). |
Repeatability | Better than ± 0.005 °C. |
Termination | Specify when ordering. |
†NVLAP Lab Code 200348-0 |