Fluke 5627A-9-X Precision Industrial RTD Probe

Fluke authorised distributor and calibration house.

The Fluke 5627a RTD temperature probe is an excellent value for a precision thermometer. It has the price-to-accuracy and price-to-durability ratios you should demand in all of your precision thermometers.

Quantity
  HTTPS Secure

Our pages are encrypted and protected from cyber-attack.

  Compliant

Payment Card Industry Data Security Standard (PCI DSS) Compliant.

  Protected

TLS (SSL) - all data/information is secure from interception.


*** Please Note: When ordering also specify the termination you require,
ie. Fluke 5627A-9-G (gold pins), plus the Bending Radius required. X = termination ***

Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “I" (INFO-CON for 1521 or 1522), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs).
All models come without calibration unless calibration is purchased separately.

The Platinum Resistance Thermometer (PRT) models 5627A is designed to be a secondary standard interpolating instrument converting temperature to resistance. The 5627A may also be used as a temperature standard for calibration of industrial sensors. The PRT is used with a readout device to detect temperature changes or actual temperature.

The Fluke 5627A-9 is manufactured using a coil suspension element design for increased shock and vibration resistance. It has a mineral-insulated sheath with a minimum bend radius of 19 mm (3/4-inch) for flexibility and durability. (Bend, if any, should be specified at time of order).

Each probe is individually calibrated and includes a NVLAP-accredited report of calibration from the manufacturer, lab code 200706-0.

Please Note: When ordering please specify the termination you require. ie. Fluke 5627A-9-G (gold pins), plus the Bending Radius required.

X = termination.

Specify “A" (INFO-CON for 914X), “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “I" (INFO-CON for 1521 or 1522), “J" (banana plugs), “L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs).

FEATURES:

Secondary PRT, 229 mm x 4.7 mm (9 in x 3/16 in), –200 °C to 300 °C.
Vibration and shock resistant.
NVLAP-accredited calibration included, lab code 200706-0.
Bending Radius: Sheath may be ordered with a bend on a minimum radius of 19 mm (3/4 in) except for 50 mm (2 in) area of sheath near tip.

Fluke 5627-9 Specifications
Resistance Nominal 100 Ω.
Temperature Coefficient 0.00385 Ω/Ω/ °C nominal.
Temperature Range −200 °C to 420 °C (5627A-6 and 5627A-9 to 300 °C; transition and cable temperature: 0 °C to 150 °C).
Drift Rate (k=2) ± 0.04 °C at 0 °C after 100 hours at 420 °C.
Sheath Material 316 Stainless Steel.
Leads PTFE-insulated, nickel-plated stranded copper, 22 AWG.
Termination Specify. See Ordering Information.
Time Constant Four seconds maximum for 63.2 % response to step change in water moving at 3 fps.
Bending Radius Sheath may be ordered with a bend on a minimum radius of 19 mm (3/4 in) except for 50 mm (2 in) area of sheath near tip. (Fluke lab requires 20 cm [8 in] of unbent sheath to re-calibrate.).
Calibration Includes manufacturer's NVLAP-accredited (lab code 200706-0) calibration and table with R vs. T values in 1 °C increments from -196 °C to 500 °C (to 300 °C for 5627A-6 and 5627A-9).
ITS-90 coefficients included.
Immersion At least 100 mm (4 in) recommended.
Calibrated Accuracy† (k=2) ± 0.026 °C at -196 °C
± 0.046 °C at 0 °C
± 0.077 °C at 200 °C
± 0.124 °C at 420 °C.
Dimensions 229 mm x 4.7 mm (9 in x 3/16 in).
†Includes calibration uncertainty and 100 hr drift.
3055528-X

Download

Fluke 5627A Datasheet

PDF User's Guide to the Fluke 5627A RTD Temperature Probe.

Download (111.79k)