loading spinner
Fluke 5606-50-X Full Immersion PRT
  • Fluke 5606-50-X Full Immersion PRT
  • Fluke 5606-50-X Full Immersion PRT
  • Fluke 5606-50-X Full Immersion PRT

Fluke 5606-50-X Full Immersion PRT

Fluke authorised distributor and calibration house.

Includes: 5606-50-X User instructions and Rigid case.

Quantity
  HTTPS Secure

Our pages are encrypted and protected from cyber-attack.

  Compliant

Payment Card Industry Data Security Standard (PCI DSS) Compliant.

  Protected

TLS (SSL) - all data/information is secure from interception.


*** Please Note: When ordering also specify the termination you require, ie. Fluke 5606-50-G (gold pins). X = termination ***

Specify “B" (bare wire), “D" (5-pin DIN for Tweener Thermometers), “G" (gold pins), “I" (INFO-CON for 1521 or 1522), “J" (banana plugs),
“L" (mini spade lugs), “M" (mini banana plugs), “P" (INFO-CON for 1523 or 1524), or “S" (spade lugs).
All models come without calibration unless calibration is purchased separately.

The Fluke 5606-50-X full immersion PRTs are designed to perform in extreme environments where both the transition junction and the lead wires are required to withstand temperatures covering the entire operating range of the probe. (“Immersion” in this case simply means exposure to temperatures other than ambient and does not necessarily refer to the medium to which the PRT and its components are subjected).

The Fluke 5606 is just two inches (50 mm) in length with a sheath diameter of 1/8 inch (3.1 mm). Since it can be fully immersed over its entire temperature range, you don’t have to worry about calculating minimum immersion depth - immerse the entire probe, transition junction, and lead wires, in either non-corrosive liquids or dry mediums ranging from –200 °C to 160 °C.

In order for any instrument to be used as a standard it must be calibrated. These instruments are sold uncalibrated unless calibration is requested at time of purchase. They are satisfactory as secondary standards and may be calibrated by comparison to primary standards.

Please Note: When ordering please specify the termination you require. ie. Fluke 5606-50-G (gold pins).

X = termination.

Specify “A” (INFO-CON for 914X), “B” (bare wire), “D” (5-pin DIN for Tweener Thermometers), “G” (gold pins), “I” (INFO-CON for 1521 or 1522), “J” (banana plugs), “L” (mini spade lugs), “M” (mini banana plugs), “P” (INFO-CON for 1523 or 1524), or “S” (spade lugs).

FEATURES:

5606: -200 °C to 160 °C.
Calibration accuracy of ± 0.05 °C.
Transition junction and lead wires can withstand full PRT temperature range.
Unique probe seal of 5606 prevents ingress of moisture.
Calibration not included; NVLAP-accredited calibration optional.

Fluke 5606 Specifications
Temperature range –200 °C to 160 °C.
Nominal resistance at 0.01 °C 100 Ω ± 0.1 Ω.
Temperature coefficient 0.00385 Ω/Ω/°C.
Accuracy[1] ± 0.04 °C at 0.01 °C
± 0.06 °C at 160 °C.
Short-term repeatability[2] ± 0.03 °C at 0.01 °C
± 0.04 °C at 160 °C.
Drift[3] ± 0.03 °C at 0.01 °C
± 0.04 °C at 160 °C.
Hysteresis ± 0.015 °C.
Sheath length 50 mm ± 5 mm, (2 in ± 0.2 in).
Sheath diameter 3.1 mm ± 0.1 mm (1/8 in ± 0.004 in).
Sheath material 316 SST.
Transition junction temperature range[4] –200 °C to 160 °C.
Sensor length 30 mm ± 3mm (1.2 in ±0.1 in).
Sensor location 3 mm ± 1 mm from tip (1.2 in ± 0.04 in).
Minimum insulation resistance 20 MΩ at 23 °C.
Minimum immersion length Full immersion.
Maximum immersion depth in liquid medium Full immersion.
Response time[5] 12 seconds typical.
Self heating (in 0 °C bath) ± 0.003 °C.
Lead-wire cable type Enameled copper wire.
Lead-wire length 2.4 meters (8 ft).
Lead-wire temperature range 160 °C.
Calibration Calibration not included. NVLAP accredited calibration available. See ordering information.
[1] Includes calibration and 100 hr drift (k = 2).
[2] Three thermal cycles from min to max temp, includes hysteresis, 95 % confidence (k = 2).
[3] After 100 hours at max temp, 95 % confidence (k = 2).
[4] Temperatures outside this range will cause irreparable damage.
[5] Per ASTM E 644.
3622827-X

Download

Fluke 5606-5607 Datasheet

PDF Specifications of the Fluke 5606 & 5607 Full Immersion PRT's.

Download (604.9k)
arrow_upward