loading spinner
Fluke 9102S Handheld Dry-Well
  • Fluke 9102S Handheld Dry-Well
  • Fluke 9102S Handheld Dry-Well
  • Fluke 9102S Handheld Dry-Well

Fluke 9102S Handheld Dry-Well

Fluke authorised distributor and calibration house.

Includes: Power cord, Fluke 9102S Dry-Well user's guide, NIST-Traceable Calibration Cert, RS-232 Cable, Interface-it Software, Insert, 3/16", Insert, 1/4" & Insert Removal Tool.

Quantity
  HTTPS Secure

Our pages are encrypted and protected from cyber-attack.

  Compliant

Payment Card Industry Data Security Standard (PCI DSS) Compliant.

  Protected

TLS (SSL) - all data/information is secure from interception.

The Fluke 9102S Mid-Range Field Calibrator is a small portable instrument designed for quick on-site checks and calibration of thermocouple and RTD temperature probes. This instrument is small enough to use in the field, and accurate enough to use in the lab. Calibrations may be done over a range of –10°C to 122°C (14°F to 252°F). Temperature display and setpoint resolution are 0.1 degrees.

The temperature is accurately controlled by Fluke’s digital controller. The controller uses a precision platinum RTD as a sensor and controls the well temperature with transistor driven thermoelectric devices.

The LED front panel shows the current well temperature. The temperature may be set, using the control buttons, to any desired temperature within the instrument’s range. Multiple fault protection devices insure user and instrument safety and protection.

This Fluke 9102S dry-well is only four inches high and six inches wide, achieves temperatures as low as -10°C, includes a NIST-traceable calibration, and is stable to ±0.05°C.

FEATURES:

A controlled temperature block with two calibration insert sleeves.
Rapid heating and cooling.
Prop stand.
Handle strap.
RS-232 interface capability.
+12 Volt DC battery option.
Temperature scan rate control.
Eight set-point memory.
Adjustable readout in °C or °F.

Fluke 9102S Specifications
Range -10 °C to 122 °C (14 °F to 252 °F) at 23 °C ambient.
Accuracy ± 0.25 °C.
Stability ± 0.05 °C.
Well-to-Well Uniformity ± 0.2 °C with sensors of similar size at equal depths within wells.
Heating times Ambient to 100 °C: 10 minutes.
Stabilisation 7 minutes.
Cooling times Ambient to 0 °C: 10 minutes.
Well depth 102 mm (4 in).
Removable inserts Available in sizes from 1.6 mm (1/16 in) to 11.1 mm (7/16 in) [6.4 mm (1/4 in) and 4.8 mm (3/16 in) included].
Power 94-234 V ac (± 10 %), 50/60 Hz, 60 W;or 12 VDC.
Computer interface RS-232 interface.
Dimensions 99 x 140 x 175 mm (3.9 x 5.5 x 6.9 in).
Weight 1.8 kg (4 lb).
Fluke 9102S warranty 1 year.